Version 4.2: Aug 26, 2015
Material residues: After cleaning, make sure that all cleaning materials have been
removed completely from the floor pan.
5. Reinstall the working plates, and then operate the workstation for at least 20 minutes at high speed so that
released aerosols can be absorbed by the filter
f. Cleaning of the humidifier
The following procedures can be applied for decontaminating the integrated humidifier. It is recommended to clean
and disinfect the humidifier monthly.
•
Turn off the gas flow.
•
Mark each tube connected to the infuser in order to recognize the inflow from the outflow.
•
Detach the tubes from the infuser. The tubes can be slightly pulled to enable this process
•
Remove and empty the bottle.
•
Fill the bottles with 2/3
rd
quaternary ammonium compounds solution.
•
Replace the bottles back into the block and connect the hoses accordingly.
•
Disconnect the external silicone hose from the glass incubator hood and place the tube in a recipient.
•
Turn on the gas flow to approximately 30l/h and leave the flow running for 20 minutes.
•
Turn the gas flow off. Disconnect the tubes and empty the remains of the bottle and rinse the bottle and
diffuser extensively with water. Empty the recipient.
•
Place the bottle, the infuser and the glass incubator hood into an autoclave.
•
Run a suitable sterilization program for the autoclave.
Please refer to the autoclave documentation provided by the manufacturer.
It is recommended to have a back-up flask and infuser as spare part to swap out or to replace
in case of damage. It is recommended to change the external silicone hoses every 6 months.
•
Once autoclaved, fill the bottle with 2/3
rd
of sterile water and put the bottle back in the humidifier assembly
at back wall of the workstation.
•
Put a small dab of mineral oil on the lid of the infuser bit to prevent it from getting stuck over time.
Make sure the mineral oil does NOT fall into the water, if this happens you will have to
clean thoroughly the bottle.
•
Connect the infuser to the hoses and place a recipient at the end of the external silicone hose.
•
Turn on the gas flow to the system at 30l/h and leave it running for 20 minutes.
•
Turn off the gas flow and empty the recipient.