X.
TROUBLESHOOTING YOUR AIR FLOW
a. Flow 1 alarm
You have an alarm on the FLOW 1. This indicates the alarm settings for the
downflow need to be adjusted.
Before starting, control that the white pre-filter mounted at the top of the
workstation are not blocked by dust or other obstructions, make sure that all
accessories, devices regularly used in the workstation are in place and not
taken out but at the same time are not blocking for the airflow passing out
through the back of the work table.
Then follow the instructions:
From STANDARD DISPLAY
Press on the RIGHT or LEFT arrow button to navigate towards the information
menu until you reach the SERVICE MODE
Press on the “ENTER” button to enter the service functions
Enter the code 1234 using the numbers found on the control panel and press
enter.
Press on the RIGHT or LEFT arrow button to navigate towards the information
menu until you reach the FLOW SENSORS menu
Press on the “ENTER” button
FLOW SENSOR 1 with as a standard type ANALOG will be displayed. If not
SWITCH will be displayed. FLOW SENSOR 1 is the down flow sensor.
Press on the “ENTER” button
NEW CALIBRATION will be displayed.
Press on the “ENTER” button