40
Version 4.2: Aug 26, 2015
IV.
Programming and Controlling the UV light timer
This section describes how to program the UV light timer.
Note: for this feature to work correctly, you must set up the date and time
first. How to do this is described in the section “Programming Time and Date”
For protection against unintended UV radiation which will harm eyes and
skin, use the timer to start the UV decontamination when no personnel is
present in the room where the workstation is located.
Standard display.
Press the “enter” button to enter the menu.
Press on the right arrow button to enter the next menu.
The USER MODE with the “UV CONTROL” will be displayed together with a
representation of the “Control & Programming Keys. The first line with the arrows
symbol “
”enables the user to go back to the previous menu or to move to
another function. The symbol “o” represents the “enter” button.
Press the “enter” button to validate your choice.
The UV AUTOSTART with START: OFF will be displayed.
Press the up or down arrow to change the settings between START: OFF, TODAY
and START: AUTO.
If you wish to program the UV light to start today, use UV AUTOSTART TODAY.
If you wish to program the UV light to start on a different day, use UV
AUTOSTART START: AUTO.