48
Anomalous time reduction and increase
Sets the required time to trigger the start and end of an anomalous aspiration
phase. Set by default to 0.3 seconds to eliminate any false microsecond
triggers. For more information on anomalous aspiration, see chapter 3.1.16.
4.3.2 Connection settings
This page consists of two sections: Active units and Available units.
When the instrument is switched on and both the Head(s) and the iPad are
correctly connected to the Bath’s WiFi network, the Bath and Head(s) should
appear in the “Available units” section. The fact that these units show up as
available means that these are recognized in the network, but not necessarily
active. To activate these units for use, they should be made active by dragging
them upwards (push/hold on the right side of the unit name) into the “Active
units” section.
For more information on connection settings in general, or for specific situations
where multiple units are to be connected within one main network, go to
chapter -- of the Operation
and Service
manual.
4.3.3 User settings
This sub menu is for creating user accounts.
To add a user, press “+”. A new user account will open up. The three tabs in
the top row show “User”, “Editor”, and “Admin”.
- An account with the “User” tab selected gives access to the Task
Menu only
- An account with the “Editor” tab selected gives access to both
the Task and the Method menu
- An account with the “Admin” tab selected gives access to all
three menus Task, Method and System.
The first created user account always has the “Admin” tab selected. Only from
the second account and on it is possible to select the desired access level.