Figure 7-28: Keratometric data
The figure 7-28 shows, as an example, the representation of the calibration sphere
with the map of the curvatures.
On the left side of the screen are shown the values of power or curvature in diopters
(Power), the radius of curvature in mm (Radius), distance from the centre in
millimetres (Distance), meridian (Meridian) and height (Height) of the point
highlighted by pointer freely placeable on the map of curvatures by pressing on the
touch screen. There are also the
Sim K
with the maximum (red) and minimum
curvature (blue) axis and the astigmatism (Diff.), the size of the pupil (Pupil) and its
decentralization (offset) from the vertex in mm and finally the E-values at 8mm.
By clicking on the
icon you can select a function directly or choose in the
information panel for the Maloney indices to be shown.