
572
Temperature Input Unit Functions
Section 11-6
Stabilization Times
required in Each Step
The following diagram shows the stabilization times (waiting times) required
when adjusting all 4 inputs.
Note
(1) The terminal block temperature stabilization time does not affect the up-
per/lower limit adjustment after changing the wiring for input 1, so the ad-
justment can be performed immediately if 30 minutes have passed since
the Temperature Input Unit's power was turned ON.
(2) To perform the adjustment procedure properly, always allow sufficient
time for the temperature to stabilize. Also allow sufficient time for devices
such as the STV, DMM, and ZERO-CON to stabilize. Refer to each de-
vice's operating manual for details.
Step
Timing chart
Stabilization time
Input Unit’s power supply ON
Wiring for input 0 adjustment
Input 0 upper/lower limit adjustment
Input 0 bias compensation
Change wiring for input 1 adjustment
Input 1 upper/lower limit adjustment
Input 1 bias compensation
Change wiring for input 2 adjustment
Input 2 upper/lower limit adjustment
Input 2 bias compensation
Change wiring for input 3 adjustment
Input 3 upper/lower limit adjustment
Input 3 bias compensation
(End of adjustment)
Wait at least 30 minutes.
(Check whether additional
time is required for devices
such as the STV, DMM,
and ZERO-CON.)
10 minutes
40 minutes
40 minutes
40 minutes
Temperature
Input Terminal’s
stabilization time
Terminal Block temperature
stabilization time 1 after wiring
Terminal Block temperature
stabilization time 2 after wiring
is changed
Terminal Block temperature
stabilization time 3 after wiring
is changed
Terminal Block temperature
stabilization time 4 after wiring
is changed
Содержание CRT1 - 10-2009
Страница 1: ...Cat No W457 E1 07 CompoNet Slave Units and Repeater Unit CRT1 Series OPERATION MANUAL...
Страница 2: ......
Страница 3: ...CRT1 Series CompoNet Slave Units and Repeater Unit Operation Manual Revised October 2009...
Страница 4: ...iv...
Страница 6: ...vi...
Страница 10: ...x...
Страница 16: ...xvi...
Страница 24: ...xxiv Conformance to EC Directives 6...
Страница 386: ...362 Units with Connectors Section 6 7...
Страница 414: ...390 Temperature Input Units Section 7 5...
Страница 430: ...406 Expansion Unit Specifications Section 8 2...
Страница 486: ...462 Bit Slave Units with Compact Connectors Section 9 5 Dimensions CRT1B MD02JS and CRT1B MD02JS 1 15 21 65 24 48 9 mm...
Страница 494: ...470 Bit Slave Units with Compact Connectors Section 9 5...
Страница 501: ...477 Repeater Unit Section 10 2 Dimensions 42 3 50 3 1 27 5 95 mm...
Страница 502: ...478 Repeater Unit Section 10 2...
Страница 508: ...484 CX Integrator Section 11 1...
Страница 509: ...485 CX Integrator Section 11 1...
Страница 602: ...578 Temperature Input Unit Functions Section 11 6...
Страница 614: ...590 Device Maintenance Section 12 3...
Страница 664: ...640 I O Power Supply Current Appendix E...
Страница 672: ...648 Index...
Страница 674: ...650 Revision History...