56
2-3. Measurement Results
1
Patient information The patient information that was entered before measurement.
2
Heart Rate/
Pulse Rate
Shows the measured heart rate. If "ECG" was set to "OFF" in the "SELECT
MEASUREMENT SENSOR" settings when the patient's information was
entered, shows the pulse rate. (refer to page 34)
3
Electrocardiogram
The ECG waveform. If "ECG" was set to "OFF" in the "SELECT
MEASUREMENT SENSOR" settings when the patient's information was
entered, this will not appear (refer to page 34).
4
Phonocardiogram
The PCG waveform. If "PCG" was set to "OFF" in the "SELECT
MEASUREMENT SENSOR" settings when the patient's information was
entered, this will not appear (refer to page 34).
5
%MAP
This value is one of the pulse waveform indices that are calculated from the
blood pressure values.
Refer to "5-10. Explanation of Technical Terms" (page 143)
6
AI
This indicates the percentage of the pressure wave effected in the carotid artery
wave that is reflected to form the reflected wave.
Refer to "5-10. Explanation of Technical Terms" (page 143); When "PCG" is
"OFF", AI is not displayed.
7
UT
Time from the start of the pulse wave to its peak. However, for measurement in
the carotid artery, the time from the start of the pulse wave to the peak of the
effected wave or the start point of the reflected wave.
Refer to "5-10. Explanation of Technical Terms" (page 143)
8
Blood pressure
value
The blood pressure values for the left and right brachium and left and right ankle.
"- - -" is printed if measurement was not possible.
9
Blood pressure left-
right difference
When the difference in the blood pressure values for the right and left brachia is
greater than 16 mmHg, the maximum blood pressure of the brachium with the
lower blood pressure is shaded.
10
Measured value
reliability
If a measured value appears in parentheses, the accuracy of that value was low
for some reason and it is indicated only for reference.
11
PWV
baPWV: The PWV value calculated from the interval from the start of the
brachial pulse wave to the start of the ankle pulse wave.
Refer to "5-10. Explanation of Technical Terms" (page 143)
12
ABI
Shows the right and left ABI values.
Refer to "5-10. Explanation of Technical Terms" (page 143)
13
Graph 1
The indices are shown on a graph. Details are explained from page 58.
14
Facility name
This is shown if a facility name was entered (refer to page 86).
15
PVR waveform
The obtained pulse wave. Because the printed amplitude is calibrated from the
measured blood pressure values, the printed amplitude may be different than
the amplitude shown on the screen.
16
Pulsatile variation
graph
This graph shows the pulsatile variation
obtained from each cuff.
a) Synchronization line: Not printed if synchro
measurement was not possible.
b) Inflation upper limit: When the inflation upper
limit is set to other than "AUTO," the setting is
printed.
c) Measurement accuracy: When "Estimate" or
"First Measurement" is displayed, accuracy may be poor.
d)
: Shows the maximum blood pressure value. Not printed if measurement
was not possible.
e) Level meter: Shows the size of the pulse wave.
f) Frame: When constriction of the brachium or ankle is suspected, the frame is
printed in bold.
17
Dr. Support
A Dr. Support is displayed based on the examination results. The content varies
depending on the level setting of "Dr. Support" in the "Print Default Settings" (page 83).
L-Ank.
Limit
BP-203RPE3(A).fm 56 ページ
9年
月
6日 土曜日 午後5時4
分
Содержание Colin BP-203RPE III
Страница 2: ...BP 203RPE3 A fm 0 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 57: ...55 2 3 Measurement Results Standard Report go to next page BP 203RPE3 A fm 55 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 150: ...148 Memo BP 203RPE3 A fm 148 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 151: ...BP 203RPE3 A fm 149 ページ 9年 月 6日 土曜日 午後5時4 分 ...