49
2-2. Basic Measurement
ECG Messages
ECG messages are explained below. When "OK" is not displayed, measurement accuracy may be
reduced. Take corrective action according to the content of the message.
Messages
Status
Corrective Action
OK
ECG is stable.
Measurement can begin.
Initializing
Initializing ECG.
Have the patient remain quiet and wait briefly.
Unstable R-R
The electrodes are dry or dirty.
Replace with new electrodes (refer to page 44).
The patient is tensing the arm, causing
myoelectric signal.
Have the patient relax the arm and rest quietly.
Noise from radio interference is
affecting the ECG waveform.
If a cell phone or other device is in use nearby,
move it farther away.
Only a weak signal can be obtained
from the wrist.
Try moving the ECG clip from the left wrist to the
instep of the left foot (secondary induction).
Check
Electrodes
An ECG electrode is not inserted in an
ECG clip.
Make sure that three electrodes are properly
inserted (refer to page 44).
The protective sheet is still affixed to
the ECG electrode.
Remove the protective sheet from the ECG
electrode (refer to page 44).
An ECG cable is not connected.
Make sure that all ECG cables are securely
connected (refer to page 128).
Note:
If "ECG" was set to "OFF" in the "SELECT MEASUREMENT SENSOR" settings when the
patient's information was entered (refer to page 34), ECG messages and the ECG wave will not
appear.
BP-203RPE3(A).fm 49 ページ
9年
月
6日 土曜日 午後5時4
分
Содержание Colin BP-203RPE III
Страница 2: ...BP 203RPE3 A fm 0 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 57: ...55 2 3 Measurement Results Standard Report go to next page BP 203RPE3 A fm 55 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 150: ...148 Memo BP 203RPE3 A fm 148 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 151: ...BP 203RPE3 A fm 149 ページ 9年 月 6日 土曜日 午後5時4 分 ...