2-1. Preparing for Measurement
6.
Attach the ECG clip for the left wrist.
• Attach the clip so that the electrode is on the inner side
of the wrist.
• Make sure that both electrodes are in full contact with
the wrist.
Attaching the PCG sensor
When necessary, wipe the application site with diluted antiseptic alcohol or a similar product.
1.
Prepare one PCG sensor pad (supplies).
2.
Remove the light blue sheet from the attachment
side (1) of the PCG sensor pad, and attach the pad
to the PCG sensor (2).
3.
Remove the transparent protective sheet from the
PCG sensor pad (3).
Caution
•The PCG sensor pad is disposable supply. Do not reuse them once they are removed. In
case that they are used on moist, injured or infected skin, dispose them right away.
Otherwise an infection may result.
•Take care that the PCG sensor does not fall on the patient. This may cause injury.
•Do not use a worn or expired PCG sensor pad. Otherwise incorrect diagnosis and
treatment may result.
(PCG sensor pad have an expiry date. After the expiry date, the pad becomes dry and
incapable of accurate measurement. Use only a pad whose indicated expiry date has
not passed. Refer to page 12 for how to confirm the expiry date.)
Notes:
• If a patient has heart murmur or abnormal sounds, the second heart sound cannot be properly
detected and measurement is not possible.
• If a patient generates noise while breathing, the second heart sound cannot be properly
detected and measurement is not possible.
BP-203RPE3(A).fm 45 ページ
9年
月
6日 土曜日 午後5時4
分
Содержание Colin BP-203RPE III
Страница 2: ...BP 203RPE3 A fm 0 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 57: ...55 2 3 Measurement Results Standard Report go to next page BP 203RPE3 A fm 55 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 150: ...148 Memo BP 203RPE3 A fm 148 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 151: ...BP 203RPE3 A fm 149 ページ 9年 月 6日 土曜日 午後5時4 分 ...