2-4. R-R Interval Examination
Use this to measure fluctuations in the interval between heartbeats in order to check the
functioning of the automatic nerve system for the cardiovascular system. The one hundred
heartbeats that are required for analysis are acquired and the resulting histogram and trend are
printed.
Starting and Ending R-R Interval Examination
1.
Prepare for measurement.
If you are performing R-R interval examination before basic measurement, go to step 2.
1. Turn the power on.
2. Enter the patient information.
Refer to "Entering and Editing Patient Information" (page 25).
3.
Attach the ECG clips to the patient.
Refer to "Attaching the ECG clips" (page 44).
2.
Make sure that "ECG: OK" is displayed on the measurement screen.
It is not necessary to attach the PCG sensor.
3.
Press the [R-R INTERVALS TEST] button
Heartbeat acquisition begins. When the effective
heartbeats count reaches 100, measurement ends and
an R-R interval exam report is printed. an R-R interval
exam report is printed.
Refer to "R-R Interval Examination Results" (page 66).
If you performed an R-R interval examination before
basic measurement, return to page 51 step 1.
4.
Remove ECG sensors from the patient.
Notes:
• R-R interval examination is only possible when ECG clips are attached.
• An R-R interval examination and basic measurement cannot be performed simultaneously.
BP-203RPE3(A).fm 65 ページ
9年
月
6日 土曜日 午後5時4
分
Содержание Colin BP-203RPE III
Страница 2: ...BP 203RPE3 A fm 0 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 57: ...55 2 3 Measurement Results Standard Report go to next page BP 203RPE3 A fm 55 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 150: ...148 Memo BP 203RPE3 A fm 148 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 151: ...BP 203RPE3 A fm 149 ページ 9年 月 6日 土曜日 午後5時4 分 ...