77
3-2. User Default Settings
HR Synchronized Tone
A beep sound can be made at each R wave of ECG during measurement. Select whether or not
this beep sound is made.
Measurement Sensors
Set default settings for attachment of sensors on the patient. Specify settings for ECG, PCG.
Measurement Sites
Set default settings for attachment of cuffs on the patient.
• Cuffs are normally attached on both brachia and both ankles.
• If a shunt is placed in an upper arm for hemodialysis, do not attach an arm cuff to or perform
measurement on that brachium.
• Measurement cannot be performed with both ankle cuffs and toe cuffs attached.
Synchro Measurement
Set the default setting for "synchro measurement".
Selections
Description
ON
Make a beep sound
OFF
Do not make a beep sound
Selections
Description
ON
Attached.
OFF
Not attached.
Selections
Description
RIGHT BRACHIUM /
LEFT BRACHIUM
ON
Arm cuff attached.
OFF
Arm cuff not attached.
RIGHT LEG /
LEFT LEG
ANKLE
Ankle cuff attached.
OFF
Ankle cuff not attached.
Selections
Description
ON
Perform synchro measurement.
OFF
Do not perform synchro measurement.
BP-203RPE3(A).fm 77 ページ
9年
月
6日 土曜日 午後5時4
分
Содержание Colin BP-203RPE III
Страница 2: ...BP 203RPE3 A fm 0 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 57: ...55 2 3 Measurement Results Standard Report go to next page BP 203RPE3 A fm 55 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 150: ...148 Memo BP 203RPE3 A fm 148 ページ 9年 月 6日 土曜日 午後5時4 分 ...
Страница 151: ...BP 203RPE3 A fm 149 ページ 9年 月 6日 土曜日 午後5時4 分 ...