51
2-2. Basic Measurement
Starting and Ending Measurement
The ECG signal and the PCG signal must be stable before measurement can begin.
1.
Make sure that "ECG: OK" and "PCG: OK" are
displayed.
• R-R interval examination can be performed before
measurement is started.
Refer to "2-4. R-R Interval Examination" (page 65).
2.
Press the [START] button on the device.
Measurement begins.
If a second measurement is required, the delay time
count is displayed. Measurement begins when the count
ends.
Note:
In some cases one measurement is required and in some cases two measurements are
required to complete measurement. Two measurements are required in the following cases:
• "Synchro measurement" setting is set to "ON" (refer to page 38).
• "Synchro measurement" setting is set to "OFF"; however, blood pressure was not correctly
measured.
• "ABI Remeasurement" setting is set to "ON" and the measured ABI is lower than the set value
(refer to page 78).