2-76
Installation Instructions
Daisy Chains
External Return Bin Installation and Service Guide
/67-2038 Rev D
© 2011 Omnicell, Inc.
c. Reconnect the 2 position yellow wire solenoid cable, and slide the internal cover over the
studs.
Figure 2-186.
Sliding RFI over studs
d. Tighten the hex nuts using the 5/16” nut driver or socket.
7. Reinstall the PCB/solenoid mount with the solenoid inserted first into the ERB housing,
clearing the return door. Keep the external return door open while reinstalling the mount
plate.
Figure 2-187.
Installing the PCB/solenoid mount
8. Maneuver the plate onto the two studs behind the nuts and position the bottom rim behind the
sheet metal tabs. Be careful not to pinch the wires behind the plate.
9. Replace the PCB/solenoid metal plate, then tighten the two nuts on either side of the plate,
using a 5/16” nut driver.
10. Close the return door.
11. Close and lock the access door with the cam keys.