50
Technical Specification
OML Remote
Classification
Class I medical electrical equipment, internally
powered, portable and hand-held, continuous
operation. No applied part(s).
Expected service life
of device
2 years
Ingress protection
rating
Rating of IP22. Vertically falling drops have no
harmful effects when the stimulator is tilted at any
angle up to 15° on either side of the vertical. Do
not get the device wet. Protected against objects
entering the stimulator of 12.5 mm and above.
Power source
Internal lithium-ion battery. Nominal voltage 3.7 V
DC. Rated for up to 1200 mAh (note capacity may
reduce with age and number of charge/discharge
cycles). Expected battery life cycle: > 500 cycles
for > 70% of initial capacity
Battery certification
Cell is approved according to UL1642 and UN38.3
certified. The battery meets the requirements
of Battery Directive and the battery packs are
RoHS compliant.
The battery is UL2054 listed and
certified according to IEC62133 edition 2.
6