45
Troubleshooting
Fault/Problem
Solution/Action
‘Not allowed’ symbol is displayed on
the OML Remote
1. This could mean that no
stimulators are joined to your OML
Remote. Join your stimulator(s) (see
page 27).
2. When at least one stimulator
is in active stimulation mode,
testing stimulation and changing
stimulation mode are not allowed.
Ensure stimulator(s) are paused
and then press Test or try to change
stimulation mode.
Stimulator(s) not responsive to the
OML Remote
1. Ensure that the stimulator(s)
are switched on and within range.
Try pressing buttons to control
stimulation again.
2. Ensure that stimulator(s) is/
are not in airplane mode (wireless
disabled).
3. Turn off stimulator and OML
Remote and try again.
4. Re-join the stimulators to the OML
Remote if the network has been lost
on them. Refer to the sections on
joining (page 27) and the stimulator
manual.
5