Electromagnetic Compliance
OML Remote
Manufacturer’s declaration of electromagnetic immunity
The OML Remote is intended for use in the electromagnetic environment
specified below. The customer or user of the OML Remote should ensure
that it is used in such an environment.
Immunity test
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD) IEC
6100-4-2
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
100 kHz repetition
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC
61000-4-5
± 2 kV
Voltage
dips, short
interruptions
and variations
IEC 61000-4-11
0% UT; 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT; 1 cycle and 70%
UT; 25/30 cycles
Single phase at 0°
0% UT; 250/300 cycle
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial, hospital or
domestic environment.
6