SAA3 3-Position Remote Safety Switch
37
OMICRON
4. Plug the serial interface connector of the
SAA3
switch into the
SAFETY
output connector of the
main device as shown in the following figure.
Figure 16: Connecting the
SAA3
3-Position Remote Safety Switch (main unit in example
CIBANO 500
)
NOTE!
Tighten the locking thumb screws of the connector firmly to ensure stable contact of the
SAA3
with
the main device.
5. Unlock the
Emergency Stop
button on the main device or test system and unlock the main device
with the key or via screen (
CIBANO 500
and
COMPANO 100
do not have a lock). If all signal lamps
glow continuously green, the main device/test system is ready for operation.
6.4
Testing
1. Unlock the
Emergency Stop
button on the main device or test system and unlock the main device
with the key or via screen
NOTE!
CIBANO 500
and
COMPANO 100
do not have a separate physical key nor a
software lock function.
2. Test the functionality of the
SAA3
3-Position Remote Safety Switch as follows:
a. Start by not pressing the
SAA3
switch (Figure 15 part I).
The main device/test system signal lamps should glow
continuously green
indicating that
all outputs are switched off.