CPC 100 V1.41
16 - 10
Table 6:
Technical data of
2kV AC
output at a
voltage range of 1000V
2kV AC output at a voltage range of 1000V
Maximum current
Peak current I
peak
< 6.0A
Effective current I
rms
< 2.5A
Output power at:
typical apparent
power S
max
typical time t
max
frequency
1000V / 2.5A
1000V / 2A
1000V / 1A
2500VA
2000VA
1000VA
60s
130s
> 2hrs
50…200Hz
50...200Hz
50...400Hz
Measurement
Voltage
Voltage and current
Range:
0 ... 1000V AC
Accuracy:
typ. error <± 0.05%(rd) ± 0.05%(fs),
guaranteed <± 0.1% (rd) ± 0.1% (fs)
Possible crest factor: < 1.5
Phase tolerance
at full scale:
typ. error < ± 0.15°,
guaranteed < ± 0.3°
Содержание CPC 100
Страница 12: ...CPC 100 V1 41 x...
Страница 28: ...CPC 100 V1 41 1 16...
Страница 90: ...CPC 100 V1 41 3 14...
Страница 194: ...CPC 100 V1 41 6 30...
Страница 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Страница 258: ...CPC 100 V1 41 9 4...
Страница 264: ...CPC 100 V1 41 10 6...
Страница 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Страница 284: ...CPC 100 V1 41 12 12...
Страница 312: ...CPC 100 V1 41 14 12...
Страница 316: ...CPC 100 V1 41 15 4...
Страница 350: ...CPC 100 V1 41 16 34...
Страница 370: ...CPC 100 V1 41 17 20 7 Finish the assembly by screwing the backnut to the housing and by tightening them with wrenches...
Страница 372: ...OMICRON Contact Addresses 22...