Others - Ramping
8 - 21
For each ramp, a trigger signal can be specified to prematurely terminate
this ramp and execute the next one. If you move to the "Trigger" cell and
press the handwheel, the cell turns into a combo box.
The following trigger conditions are available for each ramp:
• No Trigger: the ramp sweeps the selected signal from the start to the
end value within the pre-defined period of time, no trigger signal will
terminate this ramp.
• Binary: a trigger signal is fed into the Binary trigger input
Bin In
. The
ramp is terminated as soon as this signal occurs.
• IOut>: as soon as the output current exceeds a certain threshold value
(to define in column "Thresh"), the ramp is terminated.
• IOut<: as soon as the output current falls below a certain threshold
value (to define in column "Thresh"), the ramp is terminated.
• V2 AC >: as soon as the signal connected to the low level voltage
measuring input
V2AC
(0...3V AC) exceeds a certain threshold value
(to define in column "Thresh"), the ramp is terminated.
• V2 AC <: as soon as the signal connected to the low level voltage
measuring input
V2AC
(0...3V AC) falls below a certain threshold
value (to define in column "Thresh"), the ramp is terminated.
The availability of the trigger signals depends on the selected values
set at
and
of the measurement table (refer to Figure 18).
Note:
Since the determination of the r.m.s. value requires some
hundred milliseconds (
τ
≈
150ms), do not use this trigger condition for
measurements in millisecond ranges.
• Overload: the ramp terminates as soon as an Overload condition
occurs at the selected output.
What is an "Overload" trigger?
– At the
800A AC
output, an overload trigger is a condition in which
the set current cannot be reached any longer, for example,
because of an opening contact or circuit breaker.
Note:
Current values
<
50A do not initiate an "Overload" when the
current circuit opens. Therefore, if you use the trigger condition
"Overload", chose a nominal current value of
≥
50A.
9
11
12
Содержание CPC 100
Страница 12: ...CPC 100 V1 41 x...
Страница 28: ...CPC 100 V1 41 1 16...
Страница 90: ...CPC 100 V1 41 3 14...
Страница 194: ...CPC 100 V1 41 6 30...
Страница 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Страница 258: ...CPC 100 V1 41 9 4...
Страница 264: ...CPC 100 V1 41 10 6...
Страница 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Страница 284: ...CPC 100 V1 41 12 12...
Страница 312: ...CPC 100 V1 41 14 12...
Страница 316: ...CPC 100 V1 41 15 4...
Страница 350: ...CPC 100 V1 41 16 34...
Страница 370: ...CPC 100 V1 41 17 20 7 Finish the assembly by screwing the backnut to the housing and by tightening them with wrenches...
Страница 372: ...OMICRON Contact Addresses 22...