CPC 100 V1.41
8 - 20
Next, set the ramp(s) parameters, that is, the parameters that apply to this
particular ramp only. Each line of the ramps table represents one ramp.
Figure 16:
Step 2 - Set the ramp
parameters
Turn the handwheel to set the focus onto the ramps table’s first cell; in
Figure 16 above it is the output current. Then press the handwheel. The
cell turns into an entry field. Now set the output current value of your
choice by turning the handwheel, or use the number keys of
CPC 100
soft-touch keyboard. Press E
N T E R
or the handwheel to acknowledge
your setting and move to the next table cell by turning the handwheel.
Note that the units of the ramps table (the table’s columns) depend on
the selected output range.
Duration of ramp, that is, the time this particular ramp sweeps the
selected signal from the start to the end value.
After this period of time the ramp terminates and the series continues
with the next ramp (if any).
If a trigger signal occurs
before
this time has elapsed, the currently
running ramp terminates and the series continues with the next ramp. If
"SOOT" is selected, the trigger signal terminates the complete series of
ramps.
7
8
10
9
7
8
Содержание CPC 100
Страница 12: ...CPC 100 V1 41 x...
Страница 28: ...CPC 100 V1 41 1 16...
Страница 90: ...CPC 100 V1 41 3 14...
Страница 194: ...CPC 100 V1 41 6 30...
Страница 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Страница 258: ...CPC 100 V1 41 9 4...
Страница 264: ...CPC 100 V1 41 10 6...
Страница 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Страница 284: ...CPC 100 V1 41 12 12...
Страница 312: ...CPC 100 V1 41 14 12...
Страница 316: ...CPC 100 V1 41 15 4...
Страница 350: ...CPC 100 V1 41 16 34...
Страница 370: ...CPC 100 V1 41 17 20 7 Finish the assembly by screwing the backnut to the housing and by tightening them with wrenches...
Страница 372: ...OMICRON Contact Addresses 22...