![Omicron CPC 100 Скачать руководство пользователя страница 80](http://html2.mh-extra.com/html/omicron/cpc-100/cpc-100_reference-manual_3988877080.webp)
CPC 100 V1.41
3 - 4
Table 1:
Display of calculated
value in measurement
table
Measured
quantities
Display of calculated value in
measurement table
(m1) (m2)
V
V
Ratio:1, Ratio:5, Diff:
Ratio m1/m2 and phase angle
ϕ
m1 -
ϕ
m2
(if phase angles are available; else "n/a"),
and differences
Δ
U and
Δ
ϕ
U
.
I
I
Ratio:1, Ratio:5, Diff:
Ratio m1/m2 and phase angle
ϕ
m1 -
ϕ
m2
(if phase angles are available; else "n/a"),
and differences
Δ
I and
Δ
ϕ
I
.
V
I
For AC: Z or R, X
Impedance Z (magnitude in
Ω
and phase
angle
ϕ
in °) or R and X in
Ω.
For DC: R
Resistance R (in
Ω)
else "n/a"
V
I
For AC: P, Q, S
Real power P (in W) and cos
ϕ
, apparent
power S (in VA) and cos
ϕ
, reactive power
Q (in var) and cos
ϕ.
For DC: P
Real power P (in W)
else "n/a"
Содержание CPC 100
Страница 12: ...CPC 100 V1 41 x...
Страница 28: ...CPC 100 V1 41 1 16...
Страница 90: ...CPC 100 V1 41 3 14...
Страница 194: ...CPC 100 V1 41 6 30...
Страница 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Страница 258: ...CPC 100 V1 41 9 4...
Страница 264: ...CPC 100 V1 41 10 6...
Страница 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Страница 284: ...CPC 100 V1 41 12 12...
Страница 312: ...CPC 100 V1 41 14 12...
Страница 316: ...CPC 100 V1 41 15 4...
Страница 350: ...CPC 100 V1 41 16 34...
Страница 370: ...CPC 100 V1 41 17 20 7 Finish the assembly by screwing the backnut to the housing and by tightening them with wrenches...
Страница 372: ...OMICRON Contact Addresses 22...