Technical Data
47
Table 6-6:
Typical duty cycles for
operation at ambient
temperature of 23 °C
6 x 32 A (L-N)
3 x 64 A (L-N)
I
[A]
P
[W]
duty
cycle
t1
[min]
ton
[s]
toff
[s]
I
[A]
P
[W]
duty
cycle
t1
[min]
ton
[s]
toff
[s]
0 ... 25 0 ... 1200 100%
> 30 > 1800
-
0 ... 50 0 ... 1200 100%
> 30 > 1800
-
26
1400
80%
7.5
80
20
52
1400
80%
7.5
80
20
29
1300
75%
6.0
60
20
58
1300
75%
6.0
60
20
32
1200
71%
3.5
50
20
64
1200
71%
3.5
50
20
1 x 128 A (LL-LN)
I
[A]
P
[W]
duty
cycle
t1 [min]
ton
[s]
toff [s]
0 .... 80
0 ... 700
100%
> 30.0
> 1800
0
100
450
60%
4.9
30
20
120
300
43%
2.6
15
20
128
200
38%
2.0
12
20