CMC 356 Reference Manual
94
7.2.3
1 x 128 A High Current Mode (LL-LN)
1 x 0 ... 128 A (±180 A
DC
), max. 35 V
peak
, 1 x 1000 VA at 80 A
Since the current over the N socket is limited to 32 A
rms
(45 A
DC
), the third
phase is used to support the N socket. The currents 1, 2 of groups A and B
are connected in parallel.
Observe the safety instructions given in Section 7.1 on page 91 when using
this operating mode.
Figure 7-3:
Single-phase operation,
1 x 128 A high current
mode
Refer to the output curves shown in the figures 6-1 through 6-5 in section
6.3.2, "Current Outputs" on page 43.
Load
1’
N’