CMC 356-Related Products and Accessories
117
9.6.5
Adapter Cable for Scanning Heads
The adapter cable VEHK0010 connects the scanning heads
OSH256
,
TK 326
and
TVS 6.15/1
directly to a
CMC 356
test set. The scanning heads
connection cable is simply extended by the adapter cable plugging the
5-pole LEMO connectors into each other. The 16-pole LEMO connector is
plugged into the LEMO socket "ext. Interf." at the rear of a CMC test set.
From there the scanning heads are supplied with 14 VDC and meter pulses
are fed to the counter input of the CMC.
Figure 9-15:
Connecting an optical
scanning head to the
CMC 356
"ext. Interf."
connector
Figure 9-16:
Adapter cable for scanning
heads
For ordering information about scanning heads, please refer to table 9-5,
"Order numbers overview" on page 128.
!
1
3
4
VEHK0010
Optical
scanning
head
CMC 356
rear side,
connector "ext Interf."
14V
DC
Pulse input
GND
Meter to be tested