![Omicron CMC 356 Скачать руководство пользователя страница 30](http://html2.mh-extra.com/html/omicron/cmc-356/cmc-356_user-manual_3980661030.webp)
CMC 356 Reference Manual
30
5.1.1
Generator Combination Socket for VOLTAGE OUTPUT and
CURRENT OUTPUT
The combination socket CURRENT OUTPUT / VOLTAGE OUTPUT
simplifies the connection of test objects to the
CMC 356
. The three voltage
outputs (VOLTAGE OUTPUT 1-3) as well as the CURRENT OUTPUT A are
wired to the combination socket (refer to table 5-1 on page 31).
Figure 5-4:
Generator combination
socket
WARNING:
The connections on the socket are dangerous when the
test set is turned on.
Follow the safety information provided at the beginning of
this manual when connecting the generator combination
sockets.
If a dangerous voltage (greater than 42 V) is applied to the socket, a
warning indicator lights above the socket.
For currents greater than 25 A, the test object (load) should be exclusively
connected to the 4 mm/0.16 " banana sockets and not on the generator
connection socket.
Front view
View onto the connector from
the cable wiring side