CMC 356 Reference Manual
76
Figure 6-23:
Typical AC linear
progression at 50 Hz and a
sampling rate of
28.44 kHz
1
1
6.10.6.2 Channel
Cross-Talk
Conditions: sinusoidal form infeed on a channel without overload, AC
measurement on neighboring channel, integration time 1 s.
Table 6-34:
Cross talk dampening
Cross talk dampening on channels of the same potential groups in dB at
f = 50 Hz
Table 6-35:
Cross talk dampening
Cross talk dampening on channels of the same potential groups in dB at
f = 500 Hz
The cross-talk dampening on a neighboring channel of another potential
group is greater than 120 dB in all measurement ranges (f = 50 Hz or
500 Hz).
1
a)
b) 3Sigma
max
represents the maximum of the 3Sigma values of all 10 input channels.
The 3Sigma
max
value of an analog input are determined from 50 measurement values.
AC Linearity in the 100 V Range
-0.04
-0.03
-0.02
-0.01
0
0.01
0.02
0.03
0.04
0
10
20
30
40
50
60
70
80
90
100
RMS Amplitude / V
R
el
. E
rro
r /
%
3Sigmamax
Minimum-3Sigmamin
AC linearity in the 100 V range
Amplitude (rms) in V
Rel.
err
or
in
%
Actual - Expected
Full scale
x 100 %
Relative error:
Measurement range
600 V
100 V
10 V
1 V
100 mV
Dampening in dB
80
105
95
120
120
Measurement range
600 V
100 V
10 V
1 V
100 mV
Dampening in dB
65
80
75
95
95
Содержание CMC 356
Страница 1: ...CMC 356 User Manual...
Страница 38: ...CMC 356 Reference Manual 38...
Страница 100: ...CMC 356 Reference Manual 100...
Страница 104: ...CMC 356 Reference Manual 104...
Страница 132: ...CMC 356 Reference Manual 132...
Страница 136: ...The OMICRON Bootloader Software Copyright Notices 136...
Страница 138: ...Contact Information Technical Support 138...