8. Details of [Analyze] Menu
PV6253-F4E016
39
film thickness measurements in a graph. This graph is displayed as graph data when the
displayed values of film thickness measurement results are pressed with the [Add] button. The
data can be added up to a maximum of 50 points.
8.3.1. [File][Save as] menu
Saves the graph data to the film thickness history data file.
8.3.2. [File][All Clear] menu
Clears all the graph data. When the menu is selected, a confirmation message for the clearing
appears. Pressing the
[OK]
button there executes the clearing of all data. Pressing the
[Cancel]
button cancels the clearing.
8.3.3. [Set][Graph setting] menu
Displays the
[Film Thickness Graph Setup]
window, where settings for graph indication
are made.
[Automatic scale]
If this option is currently checked, the software will execute automatic scale setting in the
direction of film thickness of a film thickness graph. Every time data is added or deleted, an
appropriate maximum scale is set up, so that the data is displayed over the entire graph. The
values specified in
[Min]
and
[Max]
are not used, and the minimum scale value is always 0
[nm].
[Min]
Set the minimum scale value in the direction of film thickness of a film thickness graph
scale. If
[Automatic scale]
is currently specified, this set value will not be used. The
range for this setting is from 0 to 20000, and a difference of 100 or more from
[Max]
is
required.
[Max]
Set the maximum scale value in the direction of film thickness of a film thickness graph
scale. If
[Automatic scale]
is currently specified, this set value will not be used. The
range for this setting is from 0 to 20000, and a difference of 100 or more from
[Min]
is
required.
Содержание USPM-RUIII Series
Страница 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Страница 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Страница 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Страница 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Страница 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...
Страница 135: ...PV6670 F4E003 1 Optical Measuring Instrument Accessory Operation Manual for the U CT30 2 for the USPM RUIII Optional...