13. About the Alarm and Error Indications
PV6253-F4E016
59
13. About the Alarm and Error Indications
13.1. Errors related to communications
“An error occurred while initializing GP-IB”
The initialization of the PC card for GP-IB communication failed. Check to see if the card is
securely inserted into the slot. The program can be started up despite the occurrence of the
error, but no measurement can be conducted.
“GP-IB communication timeout error”
An error occurred in the GP-IB communication. Make certain that the power of the control
box is turned on, and that the GP-IB cables are properly connected, and then, restart the
program. The program can be started up despite the occurrence of the error, but no
measurement can be conducted.
“Communication error”
An error occurred in the USB communication or GP-IB communication. Make certain that the
power of the control box is turned on, and the GP-IB cables are properly connected, and then,
restart the program. The program can be started up despite the occurrence of an error, but no
measurement can be conducted.
13.2. Errors related to measurements
“Please lower Light intensity”
The measured data from the spectrometer is saturated, because the quantity of light is too
intense. Shorten the sampling time, and make adjustment, so that appropriate data is obtained.
Display a quantity of light graph, and operate the quantity of light monitoring button to view
the data from the spectrometer.
“Measure background”
When measuring a sample (object under measurement), the background and reference must
have priorly been measured, so that measured results can be calculated. Before measuring a
sample, measure the background.
“Measure reference”
When measuring a sample (object under measurement), the background and reference must
have priorly been measured, so that measured results can be calculated. Before measuring a
sample, measure the reference.
Содержание USPM-RUIII Series
Страница 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Страница 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Страница 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Страница 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Страница 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...
Страница 135: ...PV6670 F4E003 1 Optical Measuring Instrument Accessory Operation Manual for the U CT30 2 for the USPM RUIII Optional...