PV6088-F4E017
7
Always use the supplied power cord.
Employing any other cables than those supplied may cause failures due to
internal and/or external noises.
The supplied cables are not usable with any other devices.
When no power cord is attached to the product, refer to the
“Precautions for
proper use of the power cord
” to choose a proper power code for the unit.
Replacement of fuses
Before replacing a fuse, turn off the power and unplug the power cord so as
to avoid electric shocks.
Rating of the fuse: 250 V AC, 3.15 A
Type of the fuse: Time lag
, High breaking capacity type
Size of the fuse:
φ5.2 mm
20 mm
Replacement of the lamp
Use the specified lamp for replacement. Otherwise there is a risk of damage
or fire.
Compatible lamp: 12 V 100 W (7023 from PHILIPS)
Do not install the device in an unstable location.
If the device is installed on an unstable platform, an inclined floor, or a floor
lacking sufficient rigidity, the device could fall or collapse, which might cause
bodily injury to the user or others in the area. Prior to installation, thoroughly
check the rigidity and flatness of the place where the device is to be
installed.
Do not change the lamp outright after turning it off!
To change the lamp because it was burnt out, or for any other reason, wait
till after it has cooled down enough, because it is very hot immediately after
switched off.
Failure to conform to the following cautions may cause injury or
physical damage to peripheral devices.
Содержание USPM-RUIII Series
Страница 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Страница 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Страница 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Страница 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Страница 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...
Страница 135: ...PV6670 F4E003 1 Optical Measuring Instrument Accessory Operation Manual for the U CT30 2 for the USPM RUIII Optional...