6. Details of [Display] Menu
PV6253-F4E016
21
6.8. [Graph setting] menu
The
[Graph setting]
menu displays the
[Graph setting]
window, where detailed settings for
graphic indications of measured data are made. The settings are grouped by tabs. Make setting
by selecting the intended tab.
6.8.1. [Theory] tab
[Reference][Display reference data]
Shows the theoretical spectral reflectance data of a reflecting plane in a measured data graph
in violet.
6.8.2. [Wavelength] tab
Specify here the indication wavelength range of measured data. The specified input range will
have 380 [nm] as the minimum value on the short wavelength side, and 780 [nm] as the
maximum value on the long wavelength side, the minimum value of indication being 20 [nm].
Note: Even when the indication range width is altered, the spectral reflectance data actually
measured will be 380 to 780 [nm], and this range will be stored in the memory.
Содержание USPM-RUIII Series
Страница 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Страница 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Страница 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Страница 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Страница 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...
Страница 135: ...PV6670 F4E003 1 Optical Measuring Instrument Accessory Operation Manual for the U CT30 2 for the USPM RUIII Optional...