7. Details of [Standard] Menu
PV6253-F4E016
27
7. Details of [Standard] Menu
USPM-RU compares measured data to the preset standard and judges whether the data meets it or
not.
7.1. [Standard setting] menu
In the
[Standard setting]
menu, the standard for OK/NG judgment is set up, and the
reference values are entered. When the window is closed with the
[OK]
button, the program
judges whether the selected data meets the standard or not. NG is decided if the upper-limit
standard value is exceeded or the lower-limit standard value is undercut. The status bar at the
screen bottom shows the name of the standard file currently open.
7.1.1. Reflectance
When the
[Enable reflectance]
check box is checked, the program decides OK or NG
using the reflectance. Setting can be made in Condition 1 up to Condition 10, and the
conditions currently checked only are valid. Key in the upper-limit value and lower-limit
value. The result will be indicated as Reflectance: OK/NG in the status bar. Moreover, the
standard line will be displayed on the Main Panel.
When the
[Enable]
check box in [Average reflectance] is checked, the program decides OK
or NG using the mean reflectance. Key in the upper-limit value and lower-limit value. The
result will be indicated as Rm: OK/NG in the status bar.
When the
[Enable]
check box in [Judge: Luminous reflectance] is checked, the program
Содержание USPM-RUIII Series
Страница 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Страница 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Страница 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Страница 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Страница 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...
Страница 135: ...PV6670 F4E003 1 Optical Measuring Instrument Accessory Operation Manual for the U CT30 2 for the USPM RUIII Optional...