18 VS200 device configuration
8. Select the [
Polarization Angle Changer
] entry from the available com-
ponents and choose the [
Adjust
] entry in the [
Status
] pick list.
9. Select the [
BX3 Polarizer
] and set the Status to [
Use
].
10. Use the slide control to rotate the motorized polarizer or enter a defined
angle value in the field.
18.6.4 Create or adjust an observation method
The easiest way to create a new observation method (OM) e.g. for fluorescence
image acquisition is to copy an existing one and adjust the settings according to
your needs in the [
Device Customization
] dialog box.
Open the [Device Customization] dialog box
1. Use the [
Previously Used Layout
]
button to go to a different layout. You
can find the [
Previously Used Layout
] button at the top right in the nav-
igation bar on the VS200 ASW software's start page.
2. At the top right, on the menu bar click the [
Manual control
] button.
3. Use the [
Acquire
] > [
Devices
] > [
Device Settings
] command to open the
[
Device Settings
] dialog box.