10 Mounting of components
10
Mounting of components
10.1 Scanner
10.1.1 Objectives
The VS200 system is shipped with no objective mounted in the IX3-nosepiece.
1. Make sure that the power of the VS200 system is switched off!
2. Open the door of the VS200 scanner.
3. Make sure that the objective is clean and dust free before you mount it.
4. Depending on your VS200 kit you need to mount different objectives.
The objectives should be mounted into fixed positions as per the list below.
You can rotate the nosepiece revolver in any direction so that it is easier to
mount an objective
By default a VS200 system is shipped with a 2x and 20x objective.
Additionally recommended objectives are the 10x and 40x
UPLXAPO. Basically, all other Olympus objectives that fit into the
IX3-nosepiece can be used and are supported.
5. Remove the small plastic protection cover from the objective position.
6. Screw in the objectives counter clockwise and make sure that the objective
is screwed in tightly.
- 37 -
Instructions for use
OLYMPUS VS200 System