Shading / unneven light intensity
For specimens with very sparse samples with a lot of empty background and
rather small fluorescence emittance(e.g. FISH samples), it can occur that images
acquired with Fluorescence illumination show a variance in background illu-
mination ("shading").
If the product is being used in a network environment, note the following
points.
Make sure that the network has appropriate virus protection.
The installation of operating system service packs, security patches or third
party software updates (including runtime libraries) can cause malfunctions.
If anti-virus software is running while you are acquiring images, it can lead to
the loss of individual acquisitions. It can also slow down the acquisition pro-
cess or even cause it to be aborted altogether.
- 12 -
Instructions for use
2 Safety
OLYMPUS VS200 System