18 VS200 device configuration
18.3 Device settings - objectives
A VS200 SLIDEVIEW kit is supplied with a 2x and a 20x objective ex-works. However,
the predefined device configuration already contains a 2x, 4x, 10x, 20x and 40x
objective.
You need to delete the objectives that are not installed from the nosepiece settings
in the [
Device Settings
] dialog box.
Open the [Device Settings] dialog box
1. Use the [
Previously Used Layout
]
button to go to a different layout. You
can find the [
Previously Used Layout
] button at the top right in the nav-
igation bar on the software's start page.
2. At the top right, on the menu bar click the [
Manual control
] button.
3. Select the [
Acquire
] > [
Devices
] > [
Device Settings
] command to open the
[
Device Settings
] dialog box.
Configuring objectives
1. Select the [
IX3 nosepiece
] entry on the left side and set all objectives that
are not installed to the status [
Free
].
Beispiel
Before
After
2. To add a phase contrast objective select the [
Magnification
] entry and the
correct [
Objective Type
] entry for e.g. position 5 and add e.g. [
PH
] in the
[
Description
] field to distinguish between ‘normal’ 20x and ‘PH’ 20x.
Beispiel
- 87 -
Instructions for use
OLYMPUS VS200 System