4.3 Inspection of remaining battery power (for wireless control)
37
MAJ-2301 INSTRUCTION MANUAL
Ch.4
1
Remove the foot switch from the foot switch hanger and place it on a level, stable
surface near the surgical microscope.
Figure 4.1
NOTE
To prevent erroneous inputs, the foot switch is designed not to function when it is
tilted.
2
Turn on the surgical microscope by pressing
the power switch.
Figure 4.2
3
Press one of the switches on the foot switch
until it clicks and check the battery power
indicator and communication status indicator to
confirm that sufficient battery power is
remaining.
These indicators function as shown in
Table 4.1. If the remaining battery power is low,
replace batteries with new ones by referring to
Section 3.5, “Insertion of batteries into the foot
switch”.
Figure 4.3
Foot switch hanger
Foot switch
Power switch
Power indicator
Communication
status indicator
Battery power
indicator