3.3 Connection of wireless unit
19
MAJ-2301 INSTRUCTION MANUAL
Ch.3
3.3
Connection of wireless unit
1
After stopping the surgical microscope, check
the position of the brake pedal, push it until it
hits the back, apply the brake, and confirm that
the surgical microscope is locked to the floor.
Figure 3.1
2
Using the Allen screwdriver (3 mm width across flats), loosen the screws (× 4) on the
wireless unit attaching section of the surgical microscope and remove the cover.
Figure 3.2
CAUTION
Be sure to use the Allen screwdriver (3 mm width across flats) provided with this
instrument for work. Using a tool other than what is provided may cause equipment
damage.
NOTE
Make sure not to lose the provided Allen screwdriver (3 mm width across flats),
screws (× 4), and cover. It will be required when detaching the wireless unit from
the surgical microscope for servicing, etc.
Surgical
microscope
Brake pedal
Cover
Wireless unit attaching section
Screws (× 4)
Allen screwdriver