6.1 Care
49
MAJ-2301 INSTRUCTION MANUAL
Ch.6
Chapter 6
Care, Storage, and Disposal
WARNING
When cleaning this instrument, always wear appropriate personal protection
equipment such as eye wear, face mask, moisture-resistant clothing, and
chemical-resistant gloves that fit properly and are long enough so that your skin is
not exposed. Blood, mucus, and other potentially infectious material adhering to
this instrument could pose an infection control risk.
CAUTION
• After wiping with a piece of lint-free cloth moistened with neutral detergent, dry the
instrument thoroughly before using it again. Using the instrument while it is wet
may cause equipment damage.
• Do not attempt to clean any electrical connection part (such as a connector or
contact) of power and video cables using moistened lint-free cloth. Otherwise,
damage or corrosion may cause equipment failure.
• Do not autoclave, or gas sterilize this instrument and ancillary equipment. These
methods will damage it.
• If bioburden or infectious debris enters an area of the instrument that cannot be
cleaned, such as an opening, do not use the instrument and contact Olympus.
After using this instrument and ancillary equipment, immediately perform the following cleaning
procedures. If cleaning is delayed, residual organic debris will begin to dry, and it may be difficult to
effectively clean this instrument and ancillary equipment. Always remove debris immediately after
every use.
6.1
Care
1
Disconnect the foot switch cable from the surgical microscope and remove batteries
from the foot switch.
2
When this instrument and ancillary equipment are soiled with blood or other
potentially infectious materials, wipe off all debris using a piece of lint-free cloth
moistened with neutral detergent.
3
Remove dust, dirt, and other stains on the surface by wiping with a piece of gauze
moistened with 70% ethyl or 70% isopropyl alcohol.