42
4.5 Inspection of foot switch operation
MAJ-2301 INSTRUCTION MANUAL
Ch.4
NOTE
• The movement by the joystick is possible in one of eight directions.
• For details on the joystick operation of the foot switch, refer to the instruction
manual for the surgical microscope.
Inspecting the function of Foot switches 1 to 6
Check that the functions of the surgical microscope that are assigned to Foot switches 1 to 6
can be controlled normally.
NOTE
The default settings of Foot switches 1 to 6 of the foot switch are “Not configured”,
which means that the surgical microscope cannot be controlled unless functions
are assigned to the switches. Be sure to assign functions to the switches by
referring to the instruction manual for the surgical microscope.
Press one of the switches on the foot switch
until it clicks and confirm that the surgical
microscope function assigned to the switch is
activated normally.
Figure 4.8
Switches 1 to 6