4.4 Inspection of foot switch cable (for wired control)
39
MAJ-2301 INSTRUCTION MANUAL
Ch.4
4.4
Inspection of foot switch cable (for wired
control)
1
When batteries are in the foot switch, remove the batteries.
2
Check the appearance of the foot switch cable visually to confirm it is free of
irregularities including scratches, deformation, or peeled coating.
3
Open the terminal cover of the foot switch with your finger, hold the FOOTSWITCH
connector (on the side marked “FOOTSWITCH”) of the foot switch cable so that the
index faces upward and insert the connector into the Connection terminal on the foot
switch all the way until it clicks.
Figure 4.4
CAUTION
Be sure to turn off the surgical microscope before connecting or disconnecting the
foot switch cable. Otherwise, equipment failure or malfunction may result.
4
Hold the MICROSCOPE connector (on the side marked “MICROSCOPE”) of the foot
switch cable so that the index faces upward and insert the connector into the foot
switch terminal of the surgical microscope all the way until it clicks.
Figure 4.5
Foot switch
marking
Index
Terminal cover
Connection
terminal
Foot switch connector
Foot switch
Surgical
microscope
Index
MICROSCOPE
connector
Foot switch
cable
Foot switch terminal