32
3.6 Pairing
MAJ-2301 INSTRUCTION MANUAL
Ch.3
CAUTION
After pairing, be sure to set the pairing indicator on the foot switch and attach the
corresponding pairing sticker on the wireless unit. If you have more than one foot
switch instrument, set different pairing indicator numbers to each of them. If it is not
set correctly, unexpected functions of the surgical microscope may be activated.
NOTE
Retain the unused pairing stickers carefully so as not to lose them. They may be
necessary in case you change the pairing number.
NOTE
The communication status indicator is OFF after 2 minutes of ON. And it is OFF
after 3 seconds of flashing. To check the pairing status one more time, press a
switch on the foot switch again.
10
Using a coin or a thin object, set the pairing
indicator on the foot switch to a desired
number.
Figure 3.20
11
Attach the pairing sticker to the pairing sticker attaching section of the wireless unit.
Figure 3.21
Pairing indicator
Wireless unit
Pairing sticker