6.2 Storage
MAJ-2301 INSTRUCTION MANUAL
Ch.6
When this instrument is not to be used for a certain period, store it as instructed below. Be sure to
complete all instructions in Section 6.1, “Care” before storing this instrument.
CAUTION
• Do not store this instrument in a location exposed to direct sunlight, X-rays, radio
activity, or strong electromagnetic radiation (e.g., microwave medical treatment
equipment, short-wave medical treatment equipment, MRI, radio equipment, or
cellular phones). Damage to this instrument may result.
• When the foot switch is not to be used continuously for a long time, remove
batteries from the foot switch. Otherwise, unintended battery power consumption or
fluid leak may result.
• Before storing this instrument, confirm that its switches are not depressed.
Otherwise, the batteries will be consumed and fluid leak may also result.
4
Make sure to dry this instrument and ancillary equipment after wiping with 70% ethyl
or 70% isopropyl alcohol.
6.2
Storage
1
Remove batteries from the foot switch.
2
Store the foot switch by hanging it from the hook of the foot switch hanger or placing it
on a level, stable surface in a clean and dry place.
3
When the foot switch cable is not in use, disconnect it from the foot switch and the
foot switch terminal of the surgical microscope and store it in an immediately
accessible, clean, and dry place.
6.3
Disposal
1
Complete all instructions in Section 6.1, “Care”.
2
Dispose of this instrument and/or accessories by observing your national and local
laws and guidelines.