Chapter 6 Measurement procedures
Recording the measurement result
The image with the measurement result is recorded by pressing the [FRZ/REC]
button for at least two seconds (long push) when the measurement result is
displayed.
Exiting from measurement
Press the [LIVE/GAIN] or [VIEW] button to display exit measurement screen and
then select “OK” to exit measurement. Select “CANCEL” to return to the stereo
measurement screen.
Select “EXIT” in the stereo measurement menu to exit measuring and return to
the screen before measurement is activated.
In case accuracy cannot be achieved
(1) Bring the scope as close as possible to the object so that the screen is
filled with the object.
(2) Measure the length so that it is parallel with the screen whenever
possible. While the resolution of measuring is high when the
measuring direction is parallel with the screen, it gets low when
measuring direction becomes closer to depthwise. Therefore, the
measurement can be made more accurate by measuring a parallel
object to the screen.
(3) Change the point of view and capture an image again if the objects
have similar patterns around measurement points have few surface
characteristics or if the image is blurry or reflective.
(4) Check that the optical adapter is attached properly to the scope.
Exit measurement confirmation window
Содержание IPLEX FX IV8000-2
Страница 1: ...INSTRUCTIONS INDUSTRIAL VIDEOSCOPE IPLEX FX series ...
Страница 2: ......
Страница 10: ...2 2 Rating plate B 3 LCD monitor warning plate ...
Страница 11: ...3 4 Rating plate C 5 Scope unit warning plate 6 CARD warning mark 7 Battery warning plate 8 Cover warning mark ...
Страница 24: ...16 Chapter 2 Instrument nomenclature and functions Handle Shoulder strap Optical adapter case holder ...
Страница 26: ...18 Chapter 2 Instrument nomenclature and functions LIGHT button LIGHT indicator POWER indicator POWER button ...
Страница 203: ...195 Chapter 10 Specifications ...
Страница 206: ...198 ...
Страница 207: ......