Chapter 6 Measurement procedures
8
When the scope is set in the check tool, press the [FRZ/REC] button
to freeze the white image.
When capturing the white image, hold the scope by a position behind the
angulation section so that no force is applied to the optical adapter and
angulation section.
CAUTION
•
Applying excessive force to the optical adapter when capturing a white image may reduce
measurement accuracy.
NOTE
•
Capturing a white image
The white image affects measurement accuracy so it is necessary to adjust the brightness to
the appropriate level as shown below.
If dirt is on the white indented part of the check tool, it may not be able to capture an
adequate white image. In this case, remove the guide, which is the part with the holes for the
forward-view and side-view guides for optical adapter, from the white indented part and clean
out the dirt.
The guide can be removed by turning it counterclockwise. Use a gauze cloth with soapy
water to wipe off the dirt. To reattach the guide, repeat these steps in reverse order.
Optimum
Bright enough that the
protrusion cannot be seen
Excessive illumination
Insufficient illumination
Too bright
Protrusion
Содержание IPLEX FX IV8000-2
Страница 1: ...INSTRUCTIONS INDUSTRIAL VIDEOSCOPE IPLEX FX series ...
Страница 2: ......
Страница 10: ...2 2 Rating plate B 3 LCD monitor warning plate ...
Страница 11: ...3 4 Rating plate C 5 Scope unit warning plate 6 CARD warning mark 7 Battery warning plate 8 Cover warning mark ...
Страница 24: ...16 Chapter 2 Instrument nomenclature and functions Handle Shoulder strap Optical adapter case holder ...
Страница 26: ...18 Chapter 2 Instrument nomenclature and functions LIGHT button LIGHT indicator POWER indicator POWER button ...
Страница 203: ...195 Chapter 10 Specifications ...
Страница 206: ...198 ...
Страница 207: ......