22
10.1
- Electromagnetic Compatibility (EMC)
The Performance Table needs special attention regarding electromagnetic compatibility and must
be installed and put into use in accordance with the electromagnetic compatibility information
presented in this chapter.
Radio frequency (RF) communication equipment, portable and mobile, can affect the Performance
Table.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Performance Table is intended for use in the electromagnetic environment specified below. The customer or the user
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Grupo 1
The Performance Table uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference with nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Performance Table is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings use for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Non-applicable
Emissions due to voltage
fluctuations/scintillation
emissions
IEC 61000-3-3
Non-applicable
Guidance and Manufacturer's Declaration - Electromagnetic Emissions - I
The Performance Table is intended for use in the electromagnetic environment specified below. The customer or the user
should assure that it is used in such an environment.
IMMUNITY Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic file. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
±1 kV for input/output
lines
± 2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be
that a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 KV differential mode
± 2 KV common mode
± 1 KV differential mode
± 2 KV common mode
Mains power quality should be
that a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
the power supply
input lines
IEC 61000-4-11
<5% U
T
(dip > 95% in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(dip > 95% in U
T
)
for 5 s
<5% U
T
(dip> 95% in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(dip> 95% in U
T
) for 5
s
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Performance Table requires
continued operation during power
mains interruptions, it is
recommended that the
Performance Table be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) Magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to applications of the test level.
Содержание PERFORMANCE TABLE
Страница 1: ......
Страница 2: ......
Страница 24: ...24 11 Dimensional...
Страница 25: ...25 12 Simbology According to IEC 60601 1 and IEC 60878 standards...
Страница 35: ...35...
Страница 37: ......
Страница 38: ......
Страница 60: ...24 11 Dimensional...
Страница 61: ...25 12 Simbologia La simbolog a est en conformidad con las normas IEC 60601 1 y IEC 60878...
Страница 71: ...35...