21
10
- Technical Features
Power supply: 118/127/220/230 VC~.
A certified technician when installing the equipment must select the voltage.
Note: all equipment is adjusted at the factory to 220VC~.
Number of phases: single phase.
Frequency: 50/60 Hz.
Power 118/127V: 330 VA.
Power 220/230V: 350 VA.
Protection fuses:
−
For 220/230VC~: 2.5 A H (5 x 20 mm).
−
For 118/127VC~: 5 A H (5 x 20 mm).
−
Emergency Battery: 10 A L (6.3 x 32 mm).
−
Junction Box socket: 10 A L (6.3 x 32 mm).
Power cable specification (in compliance with requirements
6.1
and
6.2
of
IEC 60601-1-2:2010
):
−
Flexible cable PP circular 500 V 3 x 1 mm 247-5 NM 53-C5.
−
Tri-polar male plug 10 A - 250 V (
NBR 14136
).
−
Tri-polar female plug 10 A - 250 V (
IEC 60083/75
).
Electrical shock protection type:
I Class Equipment, according to
IEC 60601-1-1
and
IEC 60601-1-2
standards.
Protection degree: B Type.
Operation Mode:
−
Performance table: non-continuous operation.
Time On: 30 s. Time Off: 5 min.
−
Premium LED operating light and surgical operating light: continuous.
Premium LED operating light intensity: 8.000 to 30.000 lux.
Surgical operating light intensity: 30.000 lux.
Color temperature: 4.500 K.
Operating Environment Conditions:
−
Temperature: 15º C ~ 30º C.
−
Pressure: 75 kPa ~ 106 kPa.
−
Relative humidity: 30% ~ 70% non-condensing.
Harmful
water penetration’s protection
:
−
Equipment: IPX0.
−
Pedal: IPX1.
Thermal protection of the transformer: aperture with 130º C ± 3%.
Lifting capacity (Maximum patient weight): up to 200 kg.
Equipment Weight:
−
Net: 154,5 kg (with accessories: 170,5 kg).
−
Gross: 186,5 kg (with accessories: 259,5 kg).
Weight of the Accessories:
−
Multi-articulating headrest: 4.25 kg.
−
Headrest with clipping: 3.80 kg.
−
Knee crutches: 4,20 kg.
−
Articulating armboards: 8.20 kg.
−
Armrest for blood donation: 8.3 kg.
−
Serum support: 2.70 kg.
−
Instrument’s
tray: 5.90 kg.
−
Premium LED operating light: 7.55 kg.
−
Surgical operating light: 9.25 kg.
−
Water Unit: 2.3 kg.
−
Vacuum pump adaptor: 780g.
−
Venturi saliva ejector: 1.10 kg.
−
High power Vórtice saliva ejector: 1.2 kg.
Содержание PERFORMANCE TABLE
Страница 1: ......
Страница 2: ......
Страница 24: ...24 11 Dimensional...
Страница 25: ...25 12 Simbology According to IEC 60601 1 and IEC 60878 standards...
Страница 35: ...35...
Страница 37: ......
Страница 38: ......
Страница 60: ...24 11 Dimensional...
Страница 61: ...25 12 Simbologia La simbolog a est en conformidad con las normas IEC 60601 1 y IEC 60878...
Страница 71: ...35...