- Installation Requirements
7.1
- Pre-installation
The pre-installation must be guided by the Olsen authorized service to ensure that the environment is
adequate the specified requirements for operation of the equipment (see chapter 8) and the positioning in
which it is to be used. In this step the electrical connections and, if necessary, all the water and sewage
pipes, according to the equipment's options, must be prepared.
This equipment is not designed to be installed or operated in surgical center.
7.2
- Electrical Installation
The power grid must be single-phase and have specific grounding and 10A/30mA DR circuit breaker.
The circuit breaker must exclusively supply the Performance Table and must be easy and quick to disconnect
from the mains. If the power grid presents voltage variation, the installation of a surge protection device is
required. The power switch must be turned off at the end of the workday.
There is a table below for sizing the electrical installation.
This equipment should only be connected to a power source with protective grounding. There is risk
of electric shock!
Equipment with a junction box may present the power supply plug next to the other equipment connections,
allowing the power supply to be located inside the junction box.
7.3
- Water for the Water Unit
If your equipment has a water unit, it is essential that the water network has an easy access main
valve to stop water flow, because the water main valve should be closed daily at the end of the workday.
This valve can also be used to adjust the water pressure entering the equipment, preventing that the water
arrives with too much pressure on the water unit.
The water pressure should be between 2.8 to 4.0 bar, with ideal pH (Hydrogenionic Potential) between
6.5 to 8. If the water pressure of your hydraulic network is not enough, we recommend that you consult a
professional to review the hydraulic network.
It is recommended to use a filter before the external water supply of the equipment to prevent clogging
in the internal water system.
7.4
- Sewage System
If your equipment has a water unit, the sewage system must have good hydraulic declivity (minimum
of -2º) and should be preferably installed underground. The nominal diameter of the tubing must be of
Ø40mm.
7.5
- Compressed Air
When necessary compressed air supply, it will be necessary an oil-free dental compressor, with
dynamic pressure from 5.5 to 7.0 bar (80 to 100 PSI), and a minimum displacement of 150L/min and a 30L
tank. The use of coalescing filter in the compressed air supply is recommended.
To ensure proper air pressure, observe the following specifications for hoses according to the distance:
−
Up to 10 meters: use 1/4" hose;
−
From 10 to 20 meters: use 5/16" hose;
The use of special air compressed mesh hose is recommended. Do not use hoses for connection
between the compressor and the equipment with a distance greater than 20 meters. In these cases, it is
recommended that you consult an expert for the correct sizing of the compressor and the type of pipe being
used.
Tension (V)
Wire Diameter (mm²)
Distance (m)
Current (A)
118/127/220/230
2.5
Up to 20
5.0
Содержание PERFORMANCE TABLE
Страница 1: ......
Страница 2: ......
Страница 24: ...24 11 Dimensional...
Страница 25: ...25 12 Simbology According to IEC 60601 1 and IEC 60878 standards...
Страница 35: ...35...
Страница 37: ......
Страница 38: ......
Страница 60: ...24 11 Dimensional...
Страница 61: ...25 12 Simbologia La simbolog a est en conformidad con las normas IEC 60601 1 y IEC 60878...
Страница 71: ...35...