![OLS CASY TT Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/ols/casy-tt/casy-tt_operators-manual_3812617055.webp)
55
Troubleshooting
7.8 Control Panel Display Dark or Light
If the CASY Control Panel display is completely dark or completely light-coloured:
1
Check whether the adjustment wheel on the right side of the CASY Control Panel is out of tune.
7.9 Calibration Defective or Lost
+
If the error message CL-1 is displayed during self test, the instrument has lost its calibration par-
tially or completely (wrong check.sum).
In this case the calibration needs to be re-loaded:
1
Contact your OLS representative.
7.10 Error Message HW-3a
If the error message HW-3a is displayed during selftest, the real time clock of the system is not run-
ning. In this case:
1
Restart the real time clock as described in the operator guide (Set Date and Time in the
MAIN MENU
screen).
Control Panel Display Dark or Light