![OLS CASY TT Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/ols/casy-tt/casy-tt_operators-manual_3812617054.webp)
54
CASY TT Operator‘s Guide
7.6 Error Message PR-5 or PR-8
If the error message PR-5 or PR-8 is displayed often or even always when running a cleaning cycle
or starting a measurement, the system is not able:
+
To pump the waste from the internal fluid system into the waste container (error message PR-5).
+
To pump the fluid from the storage container into the internal fluid system (error message PR-8).
In this case try the following procedure, checking after each step to see if the problem is resolved:
1
PR-8 only: Check if there is enough CASY ton in the storage container.
2
PR-8 only: Check that the tubing going into the storage container is not kinked.
3
+
Replace the sample cup and the CASY ton storage containers with two CASY cups containing
distilled water.
+
Run three consecutive cleaning cycles.
+
Repeat these steps with freshly distilled water in the two CASY cups.
+
Empty the two cups and put them back into their position.
+
Open the
SERVICE MENU
screen and start the
Dry Liquid System
command.
+
Start the
Dry Liquid System
command a second time.
+
Put the storage container with filtered CASY ton back into its position and put a CASY cup with
freshly filtered CASY ton under the capillary.
+
Run three consecutive cleaning cycles with CASY ton.
4
In case the described steps are not successful, contact your OLS representative for maintenance
service by an OLS service technician.
7.7 Error Message PR-1a, PR-1b, PR-2a, or PR-2b
If the error message PR-1a, PR-1b, PR-2a, or PR-2b is shown during a self test after the system was
not in use for some time, one or more of the internal valves are hanging. In this case:
1
Start the system several times to run the complete selftest repeatedly. In most cases the hanging
valve will work well again after some start-ups.
2
In case the described steps are not successful, contact your OLS representative for maintenance
service by an OLS service technician.
Error Message PR-5 or PR-8