46
CASY TT Operator‘s Guide
Setting the Measuring Range and the Cursor Positions
Empty Storage container
Should the CASY System fail to completely replenish the fluid system during a
CLEAN
cycle, the
following error message appears:
ERROR SCREEN
TT-2AB-9999 15.04.2009 10:15
Error-Code: PR-8
Liquid System could not be filled
during Clean!
Please check, that there is enough
CASYton in the storage container.
Further information will be
provided by your manual.
PRESS ENTER TO CONFIRM OR PRINT TO GET A HARDCOPY
PRESS CLEAN TO FLUSH THE LIQUID SYSTEM.
Ensure that there is sufficient CASY ton in the storage container (right-hand container). Replenish
the CASY ton, if necessary, and run another
CLEAN
cycle. If the error message reappears even when
the storage container is sufficiently full, the pressure system may be malfunctioning or the internal
tube system is blocked. In this case, contact your OLS OMNI Life Science GmbH & Co. KG- instru-
ment representative.
6.5 Setting the Measuring Range and the Cursor Positions
Accurate measuring results rely on the measuring range and cursor positions being correctly
adjusted. It is particularly important for the reproducibility and comparability of the data, that all
samples to be compared are measured using identical settings. It is generally advisable to specify
the measuring range and the cursor positions for a particular cell type or a certain application after
the appropriate preliminary tests have been performed in a setup. For the required preliminary
tests, the cursors can be controlled manually using the control panel keys.
Depending on your specific application, different settings may be required. It is usually straightforward
to choose the correct setting based on the displayed size distribution. Examples for setting up major
applications such as setting up animal cell lines, bacteria and yeast can be found on www.cellcount-
ing.de. If you have any questions regarding the initial settings for your application please visit our
Online Technical Support Site at:
www.cellcounting.de
. To call, write, fax, or email us, visit the OLS
home page,
www.ols-bio.de
, and select your home country. Country-specific contact information
will be displayed.
The CASY TT settings for the measuring range and the cursors for monitoring the status of eukary-
otic cells is described below.